| Equipment
|
Maintenance Required
|
| 1. Refrigerator for storage of blood
|
Read recording temp chart frequent
at least once a day (correct temp. range
2-6oC) Test alarm system weekly. Test accuracy of alarm cut-in temp.,
once in 4 months Check counter temp inside the
cabinet with precision thermometer
periodically.
|
| 2. Deep Freeze
|
Similar to blood storage refrigerator.
Check accurate temp by thermometer
kept inside the cabinet, twice a day.
|
| 3. Bench centrifuge
|
Check accuracy of speed and time with
precision rpm meter (inchometer & stop
watch every three months
Check motor brushes at regular intervals
|
| 4. Refrigerated Centrifuge
|
Check accuracy of speed & time with
tachometer and stop watch every three
months Check temp inside the centrifuge bowl by
a temp. tester with the lid closed & head
stationary. Check motor brushes every three
months.
|
| 5. Water bath
|
Check temp 2-3 times a day Check thermometer accuracy periodically Change the water every week using only distilled water Send the discarded water for culture to rule out any fungus/microbe contaminating the water bath Use the stirrer for uniform maintenance of the temperature If no stirer is attached check temp. at diagonal end of bath periodically to ensure reasonable degree of uniformity.
|
| 6. Cell Washer
|
Check shortly before use Check speed & time monthly Check function 6 monthly Check temp. monthly Check volume dispensed by cell washer
monthly
|
| 7. Incubator
|
Check the temp. 2-3 times in a day Check accuracy of temp. periodically
|
| 8. pH meter
|
Two point calibration (check control solution pH 4-7 : 7-10) before each time of use Full maintenance every six months
|
| 9. Weighing balance
|
Check the sensitivity of electronic balances using known weights once a week Full maintenance every six months
|
| 10. ELISA reader
|
Check the reproductivility of results every 3 months Calibration graph six monthly Full maintenance every six months Check the filters for any fungal growth every 3 months Keep the filters in the dessicator
|
| 11. Autopipettes
|
Calibrate the pipettes regularly using mercury or water.
|
12. Whole blood collection equipment
- Vacuum agitator
- Non-vaccum agitator
- Balance
- Scales for weight
|
Check with known weights daily/day of use
|
| 13. Microhaernatocrit
|
Check minimum packing time once every 3 month
|
| 14. Platelet incubator
|
Check temp and agitator Check speed agitator daily
|
| 15. Thawing device
|
Check temp daily or day of use Checks for cleanliness and replace water when contaminated
|